- Manuals
- Brands
- Shimadzu Manuals
- Measuring Instruments
- UV-1800
- Instruction manual
-
Contents
-
Table of Contents
-
Bookmarks
Quick Links
Read the instruction manual thoroughly before you use the
product. Keep this instruction manual for future reference.
INSTRUCTION MANUAL
System User’s Guide
UV-1800
SHIMADZU
SPECTROPHOTOMETER
206-97040A
Feb. 2008
Related Manuals for Shimadzu UV-1800
Summary of Contents for Shimadzu UV-1800
-
Page 1
206-97040A Feb. 2008 INSTRUCTION MANUAL System User’s Guide UV-1800 SHIMADZU SPECTROPHOTOMETER Read the instruction manual thoroughly before you use the product. Keep this instruction manual for future reference. -
Page 2
This page is intentionally left blank. -
Page 3
Shimadzu representative. • To ensure safe operation, read the Safety Instructions before using the instrument. • To ensure safe operation, contact your Shimadzu representative if product installation, adjustment, or re- installation (after the product is moved) is required. -
Page 4
Emphasizes additional information that is provided to ensure the proper use of this NOTE product. The symbol used in this manual is as follows: Indication of location for related information in the instruction manuals UV-1800 SERIES… -
Page 5
Otherwise, the instrument could tip over or fall off the table. • Avoid installation sites that are exposed to corrosive gases or excessive dust. These adverse conditions may be detrimental to maintaining instrument performance and may shorten its service life. UV-1800 SERIES… -
Page 6
Installation Precautions To ensure safe operation, contact your Shimadzu representative for installation, adjustment, or re-installation after moving the instrument to a different site. WARNING • Take measures to prevent the instrument from falling in the event of earthquake or other disaster. -
Page 7
The main cover does not need to be removed for routine maintenance, inspection and adjustment. Have your Shimadzu representative perform any repairs requiring removal of the main cover. • If the power cord plug gets dusty, remove the plug from the power outlet and wipe away the dust with a dry cloth. -
Page 8
In case of electrical failure, perform the following operations: 1. Press the » » side of the power switch located on the right side bottom of the UV-1800 to turn OFF the instrument. 2. After the power comes back on, start up the UV-1800 normally by following «Installation Precautions» and «Operation Precautions». -
Page 9
Warning Labels For safety operation, warning labels are affixed where special attention is required. Should any of these labels peel off or be damaged, obtain replacements from Shimadzu Corporation. (Right side) High Temperature Light source and light source chamber are very hot. When replacing the light… -
Page 10
Our company provides a warranty on this product, as stated below. Details 1. Warranty Period: Please consult your Shimadzu representative for information about the extent of the warranty. 2. Warranty Description: If failure occurs for reasons attributable to our company during the warranty period, our company will provide repairs or the replacement of parts without charge (including USB dongles). -
Page 11
If the deuterium (D2) lamp should be broken or its life is finished, dispose of the lamp separately from general garbage. When disposing of the deuterium (D2) lamp supplied from Shimadzu Corporation, select a method which will not adversely influence the environment or human body, or ask a special disposal dealer for advice or assistance. -
Page 12
The product complies with the requirements of the EMC Directive 2004/108/EC and Low Voltage Directive 2006/95/EC. Product Name: UV-Visible Spectrophotometer Model Name: UV-1800 Manufacturer: SHIMADZU CORPORATION ANALYTICAL & MEASURING INSTRUMENTS DIVISION Address: 1, NISHINOKYO-KUWABARACHO, NAKAGYO-KU, KYOTO, 604-8511, JAPAN Authorized Representative in EU: Shimadzu Europa GmbH Address: Albert-Hahn-Strasse 6-10, 47269 Duisburg, F.R. Germany UV-1800 SERIES… -
Page 13: Table Of Contents
CONTENTS Chapter 1 General UV-1800 Configuration ……………………1-2 Components ……………………….1-3 1.2.1 UV-1800 Main Body, Front and Top Views …………….1-3 1.2.2 UV-1800 Main Body, Left Side View………………1-4 1.2.3 UV-1800 Main Body, Right Side View ……………… 1-5 1.2.4 Sample Compartment………………….1-6 1.2.5…
-
Page 14
Calibration Curve …………………… 6-14 6.3.6 Solvent Selection …………………… 6-15 6.3.7 Calibration Curve Curvature ………………..6-16 6.3.8 Spectrophotometer Types………………..6-18 Measurement System……………………6-22 6.4.1 Optical System……………………6-22 6.4.2 Electrical System …………………… 6-24 List of Cells ……………………….6-25 Cleaning the Cell ……………………..6-28 UV-1800 SERIES… -
Page 15: Chapter 1 General
1 Installation Chapter 1 General CONTENTS UV-1800 Configuration ……………………1-2 Components ……………………….1-3 UV-1800 SERIES…
-
Page 16: Uv-1800 Configuration
Fig. 1.1 Standard Contents Table 1.1 Standard Contents Check Description Part No. Qty. Spectrophotometer 206-25201-91 AC Power Cable 071-60816-12 UV-1800 Instruction Manual-System User’s 206-97040 Guide (This manual) UV-1800 Instruction Manual-Operation 206-97042 Guide UVProve Software (Install CD) 206-21439-91 UVProve Tutorial (Instruction Manual) 206-94459…
-
Page 17: Components
Components 1.2.1 UV-1800 Main Body, Front and Top Views Fig. 1.2 Front and Top Views of UV-1800 Name Description This displays the operation menus and measurement results. Keypad This is the input component for giving operation commands and numeric values to the instrument, etc.
-
Page 18: Uv-1800 Main Body, Left Side View
1.2 Components 1.2.2 UV-1800 Main Body, Left Side View 100-120, 220-240V~ 50-60H 140VA AC IN SIPPER 100-120, 220-240V~ 50-60H 140VA AC IN SIPPER Fig. 1.3 Left Side View of UV-1800 Name…
-
Page 19: Uv-1800 Main Body, Right Side View
1.2 Components 1.2.3 UV-1800 Main Body, Right Side View Fig. 1.4 Right Side View of UV-1800 Name Description USB Connector Plug in the USB memory to this connector. Power Switch Use this switch to turn ON/OFF the instrument. Press the «I» side on the switch to turn ON;…
-
Page 20: Sample Compartment
«4.3 Removing/Installing the Sample Compartment Front Cover» Multi-cell Holder Drive This is the connector for connecting the control cables for multi-cell Connector sample compartment and 8/16-cell micro multi cell holder (MMC-1600) (special accessories). UV-1800 SERIES…
-
Page 21: Keypad
Use these keys to enter numeric values. Use this key to clear a numeric value entry error. When you press this key, the numeric value which has been entered will be cleared and then you may reenter the appropriate value. UV-1800 SERIES…
-
Page 22: Light Source Compartment
The light source switch can be set anywhere in the range from 295 nm to 364 nm in 0.1 nm increments. For details, refer to the Operation Guide, «12.2 Explanation for Setting Instrument Parameters», [4. Light Source Switching Wavelength]. UV-1800 SERIES…
-
Page 23
Installation Site ……………………..2-2 Connecting Power ……………………..2-4 Checking the Light Source Lamp (D2) ………………..2-7 Operation Precautions ……………………2-8 Turning ON the Power and Initialization ……………….. 2-9 Instrument Baseline Correction ………………….. 2-13 Performance Check after Installation ………………..2-15 UV-1800 SERIES… -
Page 24: Installation Site
Installation Site 2.1.1 Installation Requirements and Preparation To use the UV-1800 properly and safely, install the instrument in an appropriate location that meets the following requirements. WARNING When using flammable and toxic samples, be sure to install ventilation equipment at installation site.
-
Page 25: Installation Space
• Ensure that there is at least 100 mm of clearance between the left side of the UV-1800 and the wall. The power supply unit and the light source unit cooling fan are located on the left side of the instrument.
-
Page 26: Connecting Power
50-60H 140VA AC IN SIPPER Fig. 2.2 The following table shows the power supply voltage, power consumption, and frequency of the UV-1800. Power Supply Voltage Power Consumption Frequency (Indication on product nameplate) 100 V — 120 V AC, 220 V — 240 V AC…
-
Page 27: Connecting To The Power Outlet
Shimadzu representative. CAUTION Verify that the power switch of the UV-1800 is OFF (i.e., «O» is pressed in) before connecting the electric plug to the outlet. Insert the connector of the power supply cable (accessory) into the power supply connector (Fig.
-
Page 28: Grounding
Ground the instrument to prevent electric shock and provide reliable performance of the UV-1800. The power supply cable supplied with the UV-1800 consists of three wires including a ground wire. When installing the instrument, be sure to insert the cable into the three-wire system outlet.
-
Page 29: Checking The Light Source Lamp (D2)
Ultraviolet light may be generated, a serious health hazard. • If the UV-1800 has been operating before opening the cover, turn OFF the instrument and let it stand until the lamp cools down sufficiently. Touching the lamp when it is still hot will burn you.
-
Page 30: Operation Precautions
• Before turning ON the power switch, check to be sure that nothing is placed in the sample compartment and cell holder. If the power is turned ON when a sample cell is mounted to the UV-1800, the light source energy check and wavelength check may be judged as «NG» since the light beam is obstructed.
-
Page 31: Turning On The Power And Initialization
Turning ON the Power and Initialization 2.5.1 Power ON/OFF Power switch Fig. 2.4 UV-1800 Power Switch Turning ON the Power Press «I» on the power switch (Fig. 2.4) to turn ON the power. The items to be initialized appear on the screen, and start being initialized or checked sequentially.
-
Page 32: Initialization Operation
This icon is displayed when the lamp is ON, and not displayed when it is OFF. A blinking icon indicates that the lamp is OFF with some errors. Fig. 2.5 Initialization screen 2-10 UV-1800 SERIES…
-
Page 33
If any abnormality is detected, however, «NG» is displayed and the initialization process is terminated. In such a case, note the item for which «NG» is displayed, and turn OFF the instrument. To learn the check points when an error occurred, refer to «5.1 Errors during Initialization». UV-1800 SERIES 2-11… -
Page 34: Switching System Language
• UV-1800series System Use’s Guide……. P/N 206-97039 • UV-1800series Operation Guide……P/N 206-97041 Switching Procedure (English to Japanese) Turn ON the UV-1800 by pressing the «I» side of the power switch (Fig. 2.4) while holding down the key on the keypad.
-
Page 35: Instrument Baseline Correction
The UV-1800 uses the correction data obtained from 1 when performing 2. Therefore, the steps and shock noises can be removed in a short time. Usually they can be removed only when corrections are performed with the same interval as that used when the measurement data is obtained.
-
Page 36
The instrument baseline correction is started. Fig. 2.9 During the correction process, a message will be displayed at the bottome of the screen. To terminate the instrument baseline correction, press the key. Fig. 2.10 2-14 UV-1800 SERIES… -
Page 37: Performance Check After Installation
• Baseline flatness • Wavelength accuracy Use the instrument validation function in order to check the performance of the UV-1800. For details on how to use the function, refer to the Operation Guide, «16. Instrument Validation Function». 2.7.1 Parameter Settings…
-
Page 38: Performing Validation
2.7 Performance Check after Installation 2.7.2 Performing Validation Connect the printer to the UV-1800, and start the validation by the following procedure: Press the [Settings] key in the Parameter Input screen for the instrument validation. Fig. 2.14 In the Settings screen, set [1.Auto Print] and [Print Init.
-
Page 39: Chapter 3 Maintenance & Inspection
3 Maintenance & Inspection Chapter 3 Maintenance & Inspection CONTENTS Inspection and Maintenance………………….3-2 Inspecting the Sample Compartment ………………..3-3 Checking and Resetting the Lamp Usage Time …………….3-4 Replacing the Light Source ………………….. 3-7 Cleaning the Exterior ……………………3-15 UV-1800 SERIES…
-
Page 40: Inspection And Maintenance
Parts». Using parts other than those specified may damage the parts, causing injuries or system malfunction. • DO NOT remove the cover of the UV-1800. Ignoring this may cause injuries or system malfunction. For internal repair of the UV-1800, contact your Shimadzu representative.
-
Page 41: Inspecting The Sample Compartment
Inspecting the Sample Compartment CAUTION DO NOT spill water or organic solvent on the UV-1800. Doing so may result in electrical failure or instrument malfunction. When analyzing liquid sample, check that no solution sample is spilled on the sample compartment before and after measurement.
-
Page 42: Checking And Resetting The Lamp Usage Time
Checking and Resetting the Lamp Usage Time The UV-1800 has a function to record and display the accumulated usage time of the WI (halogen) and D2 (deuterium) lamps used for the light source. Even though the accumulated usage time is saved to the instrument after the power is turned OFF, the data will be cleared if any electrical problem occurs.
-
Page 43: Resetting Procedure
Maintenance screen (Fig. 3.4). Fig. 3.4 The window for selecting the lamp whose usage is to be reset appears. Use the keys to move the cursor to [D2 lamp] (Fig. 3.5), and press the key. Cursor Fig. 3.5 UV-1800 SERIES…
-
Page 44
The lamp usage time will be reset. Fig. 3.6 You will return to the [Maintenance] screen (Fig. 3.7). Verify that [D2 lamp usage time] is reset to «0 hours». Press the key to return to the Mode menu screen (Fig. 3.3). Fig. 3.7 UV-1800 SERIES… -
Page 45: Replacing The Light Source
Replacing the Light Source 3.4.1 Light Source Specifications The UV-1800 uses two types of light source lamps: D2 (deuterium) lamp and WI (halogen) lamp. The D2 lamp and WI lamp are used for ultraviolet region (190 nm — light source switching wavelength and visible/near-infrared region (light source switching wavelength — 1100 nm), respectively.
-
Page 46: Lamp Replacement Procedure
• After inserting the WI lamp into the socket, DO NOT forcedly move it to right and left or up and down. The connection part between the pin and glass may crack, which makes it impossible for the lamp to illuminate. UV-1800 SERIES…
-
Page 47
Fixing screw While sliding the light source compartment cover and lifting it up at an angled (following the arrow direction in Fig. 3.9), remove it from the main body. Fig. 3.9 Removing the light source compartment cover UV-1800 SERIES… -
Page 48
D2 lamp before the work. To remove the D2 lamp, refer to » Replacing the D2 lamp» in this section. Remove the WI lamp retainer spring from the top of the WI lamp. Fig. 3.11 3-10 UV-1800 SERIES… -
Page 49
Install the light source compartment cover by the reverse procedure of that described in » Removing the light source compartment cover» in this section. Insert the electric plug into the outlet, and switch ON the UV-1800. (Press the «|» side on the switch.) UV-1800 SERIES 3-11… -
Page 50
(i.e., «OK» is displayed for all items.) When the Mode menu screen is displayed, follow the procedure in «3.3.2 Resetting Procedure» to reset the WI lamp usage time. Fig. 3.16 3-12 UV-1800 SERIES… -
Page 51
Notch (on the socket) Locating lug Fig. 3.18 Insert the electric plug into the outlet, and switch ON the UV-1800. (Press the «|» side on the switch.) Fig. 3.19 UV-1800 SERIES 3-13… -
Page 52
(i.e., «OK» is displayed for all items.) When the Mode menu screen is displayed, follow the procedure in «3.3.2 Resetting Procedure» to reset the D2 lamp usage time. Fig. 3.20 3-14 UV-1800 SERIES… -
Page 53: Cleaning The Exterior
Then, wipe the moisture off with a dry cloth. NOTE DO NOT leave any spilled water on the UV-1800. DO NOT use alcohol or paint thinner solvents for cleaning. Doing so may cause the instrument surface to rust or discolor.
-
Page 54
3.5 Cleaning the Exterior This page is intentionally left blank. 3-16 UV-1800 SERIES… -
Page 55: Chapter 4 Replacing The Sample Compartment Parts
4 Replacing the Sample Compartment Parts Chapter 4 Replacing the Sample Compartment Parts CONTENTS Removing/Installing the Cell Holder………………..4-2 Removing/Installing the Sample Compartment Unit (Standard) …………4-4 Removing/Installing the Sample Compartment Front Cover …………. 4-7 UV-1800 SERIES…
-
Page 56: Removing/Installing The Cell Holder
Standard cell holder Knurled screw Fig. 4.1 Sample compartment (standard) 4.1.1 Removing the Cell Holder Open the sample compartment cover. Loosen the knurled screw fixing the cell holder. Remove the cell holder. Knurled screw Cell holder Fig. 4.2 UV-1800 SERIES…
-
Page 57: Installing The Cell Holder
NOTE Install the cell holder so that the beam passes through it as directed by the arrow mark. Cell holder Fix the cell holder with the knurled screw. Locating pin Locating hole Fig. 4.3 UV-1800 SERIES…
-
Page 58: Removing/Installing The Sample Compartment Unit (Standard)
Fig. 4.4 Main body front view 4.2.1 Removing the Sample Compartment Unit Loosen two of the sample compartment unit fixing Fixing pin (on the UV-1800) screws located at the bottom of the sample Notch (on the sample compartment.
-
Page 59: Installing The Sample Compartment Unit
Open the sample compartment cover and install the sample compartment unit. At an angle from above, insert the notch on Locating pin Notch (on the sample (on the UV-1800) the sample compartment unit to the locating compartment unit) pin at the far side of the sample compartment.
-
Page 60
4.2 Removing/Installing the Sample Compartment Unit (Standard) Make sure that the sample compartment unit front cover and the UV-1800 main body fit together snugly. If they don’t, press the sample compartment unit down to the sample compartment bottom. Fig. 4.9… -
Page 61: Removing/Installing The Sample Compartment
Fig. 4.11 Front plate for syringe sipper (switching unit) 4.3.1 Removing the Sample Compartment Front Cover and Installing the Front Plate Remove the sample compartment unit from the UV-1800 by the procedure described in «4.2.1 Removing the Sample Compartment Unit».
-
Page 62: Installing The Sample Compartment
4.3 Removing/Installing the Sample Compartment Front Cover Install the sample compartment unit (without cover) to the UV-1800 main body by the procedure described in «4.2.2 Installing the Sample Compartment Unit». Install the designated front plate (special accessory) to the sample compartment unit.
-
Page 63
Snap the sample compartment front cover into the sample compartment unit. Push the cover in the arrow direction in Fig. 4.17 until it snaps. Install the sample compartment unit to the UV-1800 main body by the procedure described in «4.2.2 Installing the Sample Compartment Unit». -
Page 64
4.3 Removing/Installing the Sample Compartment Front Cover This page is intentionally left blank. 4-10 UV-1800 SERIES… -
Page 65
5 Troubleshooting Chapter 5 Troubleshooting CONTENTS Errors during Initialization ……………………5-2 Problems: Symptoms and Solutions………………..5-4 UV-1800 SERIES… -
Page 66: Errors During Initialization
Errors during Initialization After the power is turned ON, the UV-1800 initializes and checks the following items listed in the table below. Fig. 5.1 Initialization screen Table 5.1 Initialization items Item No. Initialization items Description LSI Initialize Initializes each I/O device.
-
Page 67
5.1 Errors during Initialization When [NG] is displayed during initialization, perform the appropriate remedial actions by referring to the table below. If the problem is not resolved by these actions, contact your Shimadzu representative. Reference Item No. Check Point Remedial Action… -
Page 68: Problems: Symptoms And Solutions
Problems: Symptoms and Solutions Check whether the problem exhibits the following symptoms before requesting a repair. Contact your Shimadzu representative if the error cannot be resolved through the remedial action described below, or if the symptom is not covered in the table.
-
Page 69
The lamp is currently OFF with some errors. The fan is not working or the thermo —— censor on the circuit board has detected overheating. Turn OFF the UV-1800 power switch and contact your Shimadzu representative. Did you mistakenly Return to the blank condition press the… -
Page 70
«Cannot load this file.» Has the file been The files once saved in the UVProbe saved in the cannot be loaded on the UV-1800. —— UVProbe? Has the file been The selected file may be corrupted. properly saved? In that case, the file cannot be ——… -
Page 71: Chapter 6 Reference
6 Reference Chapter 6 Reference CONTENTS Specifications ………………………. 6-2 Maintenance Parts……………………..6-8 Spectrophotometer Basics …………………… 6-9 Measurement System……………………6-22 List of Cells ……………………….6-25 Cleaning the Cell ……………………..6-28 UV-1800 SERIES…
-
Page 72: Specifications
15 °C ~ 35 °C Operating Humidity 30 % ~ 80 % (No condensation, 70 % or less at 30 °C or higher) Power Supply AC 100 V/120 V/220 V/230 V/240 V, 50/60 Hz Power Consumption 140 VA UV-1800 SERIES…
-
Page 73: Software Specifications
— 0.001 to 0.001) %T, E; -400.0 to 400.0 (Min. range; — 0.1 to 0.1) (5) Scan repetitions: 1 to 99 (6) Recording method: Overlay/ Sequential selectable (7) Reduce/Expand [zoom], Display data with cursor (8) Save/Load data function UV-1800 SERIES…
-
Page 74
(3) Measurement using Abs difference between two wavelengths is possible. Rate-measurement (1) Measure Abs time change and calculate the change in absorbance quautity. (2) Measurement time: 1 to 9999 sec/min (3) Determine whether the measured reaction result is progressing in linear fashion. UV-1800 SERIES… -
Page 75
DNA concentration =K1×A1-K2×A2 Protein concentration = K3×A2-K4×A1 (2) Factor/measurement wavelength can be set arbitrarily. (3) Background correction can be applied. Protein quantitation (1) Quantitation method: Lowry method, BCA method, Biuret method, CBB method (Bradford method), UV absorption method UV-1800 SERIES… -
Page 76
• Peak/Valley detection (both possible up to 20 pcs.) • The four rules of arithmetic • Derivative processing • Smoothing processing • Area calculation • Point-pick processing • Print processing (printing waveform data in text format) UV-1800 SERIES… -
Page 77
: Max. 8 files (4) Accumulation time set function (for fixed wavelength measurement) (5) External control function UV-1800 can be controlled from PC. This function is applied for the control with bundled UVProve software. * USB cable (that supports USB 1.1) separately required… -
Page 78: Maintenance Parts
Fuse, 218 004 072-02004-22 For 100-120 V, 220-240 V Battery, CR2032H 38 074-73306-08 O-ring 036-15501-21 Used for fixing the quartz plate. Alternate Sample Compartment 206-60184-07 Standard cell holder 206-82009-91 Sample compartment front cover 206-25269 ASSY Sample compartment cover 206-25269 UV-1800 SERIES…
-
Page 79: Spectrophotometer Basics
Ultraviolet rays (vacuum ultraviolet) 50 nm X-rays 5 pm γ-ray (0.05 Å) (Note) 1 mm = 10 m, 1 µm = 10 m, 1 nm = 10 m, 1 pm = 10 Fig. 6.1 Types of electromagnetic waves UV-1800 SERIES…
-
Page 80
580 — 595 Yellow Blue 480 — 490 Greenish blue Orange 595 — 610 Orange Greenish blue 490 — 500 Bluish green 610 — 680 Bluish green 500 — 560 Green Purplish red 680 — 700 Purplish red Green 6-10 UV-1800 SERIES… -
Page 81: Ultraviolet/Visible Spectrum
In addition, the molecule may emit the absorbed light energy as light when returning to the ground state. Fluorescent light and phosphorescence exhibit the behavior and analysis utilizing these phenomena is called fluorometry. Excited state 1 Excited state 2 Light Ground state Fig. 6.4 Molecule energy UV-1800 SERIES 6-11…
-
Page 82: Bouguer-Beer’s Law
ε. This molar absorptivity becomes a value unique to the substance under specific conditions. To fulfil Bouguer-Beer’s law, it is necessary to satisfy conditions such as being free from stray light, emission, scattering, and reflection. 6-12 UV-1800 SERIES…
-
Page 83: Qualitative Analysis And Quantitative Analysis
Analysis to perform quantitative analysis by comparing the color darkness of a substance is called colorimetric analysis. When the substance is transparent, if absorbance exists in the invisible ultraviolet/near infrared area, it is measured. The latter is broadly included in colorimetric analysis. UV-1800 SERIES 6-13…
-
Page 84: Calibration Curve
This method is applied only when the related curve is straight as far as to the low concentration range. Generally, extra-large absorbance wavelength is used as measurement wavelength for quantitative analysis. 6-14 UV-1800 SERIES…
-
Page 85: Solvent Selection
Dichloromethane Chloroform Metylacetate Metyl formic acid Carbon tetrachloride Dimethyl sulfoxide Benzene Xylene Toluene Ethylene tetrachloride Acetone Bromoform Nitro methane Carbon disulfide 450 (nm) Fig. 6.7 Wavelength range for solvent to be used (using a 10 mm cell) UV-1800 SERIES 6-15…
-
Page 86: Calibration Curve Curvature
0.0211 0.1791 0.002 0.0000 0.0000 0.0001 0.0003 0.0005 0.0008 0.0027 0.0085 0.0791 0.001 0.0000 0.0000 0.0000 0.0001 0.0002 0.0004 0.0013 0.0043 0.0413 0.0001 0.0000 0.0000 0.0000 0.0000 0.0000 0.0000 0.0001 0.0004 0.0043 (Test and Engineering Vol.9 No.9,702,1981) 6-16 UV-1800 SERIES…
-
Page 87
If the bandwidth is made large, the peak height becomes low in appearance and this may cause the calibration curve to be bent. Spectrum band width Real spectrum half-width Fig. 6.9 Band width and peak height UV-1800 SERIES 6-17… -
Page 88: Spectrophotometer Types
I and the reference sign Io simultaneously. Reference side Cell Light source Monochromator Detector Amplifier Display Sample side (Rotating mirror, semi- Sample compartment transparent mirror) Fig. 6.11 Double beam configuration 6-18 UV-1800 SERIES…
-
Page 89
Deuterium lamp Window plate S1-2: Slit L.M.: Light-source switching mirror M1-5: Mirror Grating Filter B.S.: Beam splitter Sam: Sample side cell Ref: Reference side cell Lens P.D.: Photodiode Fig. 6.13 Double beam — double detector configuration (UV-1800) UV-1800 SERIES 6-19… -
Page 90
Halogen lamp Deuterium lamp Window plate S1-3: Slit M1-12: Mirror G1-2: Grating Filter Chopper mirror Photomultiplier Fig. 6.14 Double monochromator configuration (UV-2500 series) 6-20 UV-1800 SERIES… -
Page 91
Sample side cell (First monochromator) W1 ~ W3: Window plate (ø 30 mm) G3, G4: Diffraction grating W4 ~ W5: Window plate (ø 40 mm) (Second monochromator) C.H.: Chopper mirror Fig. 6.15 Double monochromator configuration (UV-3600 series) UV-1800 SERIES 6-21… -
Page 92: Measurement System
Measurement System 6.4.1 Optical System A schematic of the optical system for the UV-1800 is shown in Fig. 6.16. L.M.: Light-source switching mirror Deuterium lamp M1~M5: Mirrors Halogen lamp B.S.: Beam splitter Filter Lens Grating Sam: Sample cell Entrance slit…
-
Page 93
The monochromator slit aperture is fixed at 1 nm. The Shimadzu blazed holographic grading is used for the diffraction grading. The light coming out of the monochromator passes through a stray light cutting filter F and strikes the mirror M3 and is then split by the beam splitter B.S. -
Page 94: Electrical System
6.4 Measurement System 6.4.2 Electrical System A schematic of the electrical system for the UV-1800 is shown in Fig. 6.18. The center of control is the microcomputer (CPU), which performs all controls of light sources, switching of light sources, filter switching, wavelength scanning pulse motor, LCD monitor display, keyboard and printer, USB interface etc.
-
Page 95: List Of Cells
Micro multi-cell (8 cells) optical length 10 mm and holder req’d 208-92089 required sample volume of 100 µl more (206-23680-91 206-23690-91) Micro multi-cell Micro multi-cell (16 cells) optical length 10 mm holder req’d 208-92088 and required sample volume of 100 µl more (206-23680-91 206-23690-91) UV-1800 SERIES 6-25…
-
Page 96
Standard required sample volume: 1.0 ml or more Square flow cell (semi-micro) for syringe sipper, light path length: 10 mm 208-92005 390 µl 3.5 × 11 Unnecessary With tube Standard required sample volume: 5.0 ml or more 6-26 UV-1800 SERIES… -
Page 97
6.5 List of Cells 9 mm pitch 4.5 mm pitch unit : mm Fig. 6.19 Optional Cell Shapes UV-1800 SERIES 6-27… -
Page 98: Cleaning The Cell
«PVC tube for peristaltic pump» since the tube corrodes and hardens from the solvent. Be sure to fill the flow cell with distilled water after washing. If the flow cell is left empty, the cell may get stained and it becomes hard to clean afterwards. 6-28 UV-1800 SERIES…
-
Page 99
Record of Revision Date Revision Changed Page Description 2/2008 Front, 2-15, 6-2, 6-8, and more A: Record of Revision Note) A …Added Page No. D …Deleted Page No.
- Описание
- Назначение
- Преимущества
- Комплектация
- Технические характеристики
- Поддержка
- Отзывы (0)
Спектрофотометр УФ-1800 (Эковью) с первичной поверкой
Спектрофотометры УФ-1100, В-1100, УФ-1200, В-1200, УФ-1800, УФ-3000, УФ-3100, УФ-3200, УФ-6100 предназначены для измерения коэффициента пропускания, оптической плотности и концентрации веществ в жидких пробах.
Назначение спектрофотометра УФ-1800:
Внедренные инновационные технологи:
- Сенсорный дисплей.
- Возможность сохранения результатов на внешнее устройство хранения данных.
- Возможность переносить градуировку между однотипными приборами.
- Возможность ежедневно сохранять результаты измерений в памяти прибора и просматривать их в любое время.
Преимущества спектрофотометра УФ-1800:
- Отличные технические характеристики оптической схемы позволяют решать с высокой точностью любые аналитические задачи.
- Система подсказок оператору позволяет работать с прибором даже неопытному пользователю.
- Оптическая схема с автоматической установкой длины волны обеспечивает высокую надежность прибора и непревзойденную точность измерений.
- Интуитивно-понятный пользовательский интерфейс и сенсорный дисплей позволяют значительно упростить работу Оператора и уменьшить затраты времени на создание градуировки.
- Привычные режимы работы обеспечивают быстрое проведение рутинных анализов.
- Кюветное отделение позволяет использовать любые типы кювет с длиной оптического пути до 100 мм.
- Внутренний журнал позволяет сохранять результаты измерений и впоследствии их просматривать.
- Возможность переноса данных с использованием USB-накопителей позволяет использовать одну градуировку для однотипных приборов в лаборатории.
- Система автоматической юстировки длины волны позволяет освободить Пользователя от постоянного контроля точности прибора при помощи светофильтров.
- Современные электронные компоненты гарантируют длительный срок эксплуатации.
- Возможность подключения автоматического 8-ми позиционного кюветодержателя и широкого перечня дополнительного оборудования позволяют уменьшить затраты времени Пользователя при решении аналитических задач.
Комплект поставки спектрофотометра УФ-1800:
- Спектрофотометр — 1 шт.
- Держатель для 3-х кювет КФК до 100 мм- 1 шт.
- Шнур питания — 1 шт.
- Чехол от пыли — 1 шт.
- Лампа галогенная запасная — 1 шт.
- Заглушка-адаптер — 3 шт.
- Кюветы К-8 10 мм. КФК — 4 шт.
- Кюветы КУ-1 10 мм. КФК — 2 шт.
- Кабель для подключения к ПК -1 шт.
- Диск с ПО для подключения к ПК — 1 шт.
- Копия Свидетельства об утверждении типа СИ — 1 экз.
- Копия Сертификата соответствия — 1 экз.
- Копия Описания типа СИ — 1 экз.
- Копия Методики поверки — 1 экз.
- Паспорт — 1 экз.
- Свидетельство о поверке — 1 экз.
- Руководство по эксплуатации — 1 экз.
- Руководство пользователя к ПО — 1 экз.
Технические характеристики спектрофотометра УФ-1800:
Спектральный диапазон, нм | от 190 до 1100 |
Диапазон измерений спектральных коэффициентов направленного пропускания, % | от 0,1 до 99 |
Диапазон показаний спектральных коэффициентов направленного пропускания, % | от 0 до 200 |
Диапазон показаний оптической плотности, Б | от -0,3 до 3,0 |
Пределы допускаемой абсолютной погрешности спектрофотометра при измерении коэффициентов направленного пропускания, %: — в спектральном диапазоне от 400 до 800 нм — в остальном спектральном диапазоне |
±0,5 ±1,0 |
Пределы допускаемой абсолютной погрешности установки длин волн, нм | ±0,5 |
Спектральная ширина щели, нм | 2,0 |
Дрейф показаний, Б/ч, не более | ±0,002 |
Отклонение нулевой линии от среднего значения (в диапазоне от 300 до 800 нм), Б, не более | ±0,002 |
Уровень рассеянного света (при λ=340 нм), %, не более | 0,05 |
Габаритные размеры (ДxШxВ), мм | 450×380×180 |
Масса, кг, не более | 12 |
Потребляемая мощность, В•А, не более | 140 |
Средний срок службы, лет | 8 |
Наработка на отказ, ч, не менее | 6400 |
Напряжение питания частотой (50±1) Гц, В | 220+22-33 |
Условия эксплуатации — диапазон температур окружающего воздуха, ºC — диапазон относительной влажности окружающего воздуха (при 25ºС), % — диапазон атмосферного давления, кПа |
от 15 до 30 не более 80 от 84 до 106 |
В наличии документация на медицинское оборудование фирмы Shimadzu (в т.ч. на русском языке):
Инструкция по эксплуатации (Operation (Instruction) manual) на Digital Radiographi System SDR-100 [Shimadzu] Рентген
Инструкция по монтажу (Installation instructions) на Digital Radiographi System SDR-100 [Shimadzu] Рентген
Инструкция по эксплуатации (Operation (Instruction) manual) на FlexaVision [Shimadzu] Рентген
Инструкция оператора (Operator manual) на FlexaVision R/F System [Shimadzu] Рентген
Инструкция по установке и обслуживанию (Servise and Installation manual) на FlexaVision R/F System [Shimadzu] Рентген
Сервисная инструкция (Service manual) на Mobile DaRt MUX-100D [Shimadzu] Рентген
Техническая документация (Technical Documentation/Manual) на Mobile X-ray System: MUX-100 [Shimadzu] Рентген
Инструкция оператора (Operator manual) на Mobile X-ray System: MUX-100 [Shimadzu] Рентген
Руководство по установке (Installation Manual) на Mobile X-ray System: MUX-100 [Shimadzu] Рентген
Схема электрическая (Electric scheme (circuit)) на RADspeed Fit [Shimadzu] Рентген
Руководство пользователя (Users guide) на SDU-1200, SDU-2200 [Shimadzu] Диагностика-УЗИ
Схема электрическая (Electric scheme (circuit)) на X-Ray Computerized Tomography system SCT 7800 [Shimadzu] Рентген
Инструкция по эксплуатации (Operation (Instruction) manual) на X-Ray Computerized Tomography system SCT 7800 [Shimadzu] Рентген
Инструкция по монтажу (Installation instructions) на X-Ray radiography table BK-12H/BK-12HK [Shimadzu] Рентген
Инструкция оператора (Operator manual) на ZSU-5D/5DS Remote Controlled R/F System [Shimadzu] Рентген
Инструкция по монтажу (Installation instructions) на ZSU-5D/5DS Remote Controlled R/F System [Shimadzu] Рентген
Инструкция по монтажу (Installation instructions) на ZUD-P40D, L40D, L41D, V40D, B40D [Shimadzu] Рентген-Генератор
Методические материалы (Methodical materials) на ПО UVProbe Обучение Версия 1.1 [Shimadzu] Лаборатория
Методика поверки (Methods of verification) на Спектрофотометр UV серии [Shimadzu] Анализаторы-Фотометр
Инструкция по эксплуатации (Operation (Instruction) manual) на Спектрофотометр UV-1800 [Shimadzu] Анализаторы-Фотометр
Хранилище файлов |
Правила добавления:
Внимание !!!
Перед тем как загрузить файл, убедитесь, что его нет в «Хранилище файлов» для этого воспользуйтесь поиском.
При добавлении файлов в поле «Название материала» указываем марку аппарата, так как на шильдике или по паспорту.
В поле «Краткое описание» пишем что-это: схема; сервис мануал; руководство по эксплуатации и т.д. Также напишите формат материала, какой программой он сделан.
За сохранность и работоспособность файлов(ссылок) загруженных на сторонние файловые хостинги, Администрация сайта ответственности не несет.
Для загрузки файлов в «Хранилище» воспользуйтесь архиватором WinRAR или 7-Zip, максимальный объем загружаемого файла должен быть не более 15Mb.
Администрация сайта оставляет за собой право удалять и редактировать файлы.
Для скачивания с 4Shared.com необходимо зарегистрироваться и создать аккаунт на 4shared.com или пользоваться другим способом скачивания которые там же предлагаются.
Спектрофотометр Shimadzu UV-1800
Гостям запрещено просматривать данную страницу, пожалуйста войдите на сайт как пользователь.
[
Регистрация
|
Вход
]
14.Апр.2013, 18:00 | |
Категория: Документация: | |
Просмотров: 3359 | Загрузок: 65 | Размер: 8.57 Mb
|
Вопросы по ремонту медицинской техники | Теоретические основы медицинского приборостроения | Информационный раздел | Информация о сайте | Хранилище файлов | Сертификаты и регистрационные удостоверения | Навигатор по сайту |
Написать администратору
Спектрофотометр Shimadzu UV-1800 – это современный двухлучевой спектрофотометр, позволяющий проводить любые измерения количественного анализа жидких и твердых веществ. UV-1800 – один из лидеров рынка, производится в Японии, но полностью соответствует европейским, российским и азиатским требованиям эксплуатации.
Спектрофотометр Shimadzu UV-1800 характеризуется высокой точностью производимых измерений, отличными оптическими характеристиками, надежностью в использовании, простотой в ознакомлении и дальнейшей работе с прибором, современным дизайном. UV-1800 имеет очень компактные размеры, его ширина всего 270 мм, он поместится на любом столе и не займет много места, но всегда окажется под рукой.
Результаты анализа отображаются на жидкокристаллическом экране. Данные можно передавать на ПК с помощью USB. Кроме того, с помощью программного обеспечения UVProbe, установленного на ПК, можно все управление спектрофотометром вести непосредственно через компьютер. Также можно подключать принтер и сразу распечатать результаты исследований. Просто и удобно.
Shimadzu UV-1800 согласно встроенным программам способен работать в нескольких режимах: фотометрический, спектральный, кинетический, количественный. Кюветное отделение очень удобное – кюветы меняются без особых усилий.
Не удивительно, что ввиду своих характеристик спектрофотометр Shimadzu UV-1800 получил широкое распространение и сегодня успешно работает в химических и научных лабораториях, на фармацевтических предприятиях, в санитарных и контролирующих учреждениях.
UV-Vis Spectroscopy
Maintaining the highest level of wavelength resolution in its class (1 nm), Shimadzu’s latest UV-VIS spectrophotometer, the UV-1800, can function as a standalone instrument or can be PC-controlled with USB memory.
UVProbe data processing software now comes standard so users can drag and drop data from an optional USB memory stick onto a PC. The USB interface gives users the advantage of macros and graph functions available with commercial spreadsheet software to reduce effort and enhance efficiency. USB memory saves spectral data, quantitation values and measurement parameters. In addition, it saves parameter settings so multiple users can share one instrument.
With the UV-1800, stray light and noise levels are reduced. Shimadzu’s LO-RAY-LIGHT low stray light diffraction grating restricts stray light to half that of the UV-1700. A brighter optical system reduces the noise level (at 700 nm) to 25 percent of the previous model.
Multiple modes, including photometric, spectrum, quantitation, kinetics, time-scan, multi-component quantitation and biomethod, come standard on the UV-1800. The UV-1800 also offers automated security, and enhanced instrument validation and maintenance functions.
The new design includes a 15-percent smaller footprint (45 cm W x 49 cm D) than its predecessor for placement close to other equipment; fewer keys that are spaced farther apart for trouble-free operation; and three display screens — parameter settings, progress status and results display — for clear information and ease of use.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Manuals & DocumentsFeaturesSpecifications
Documents & Manuals
There are no Documents or Manuals available.
Features of UV-1800
High resolution – 1 nm
Featuring the highest resolution in its class*, the UV-1800 easily satisfies the standards of wavelength resolution demanded by the European Pharmacopoeia. Additionally, using a spectroscope with a Czerny-Turner mounting results in a compact, bright optical system. Stray light, wavelength repeatability and baseline stability have also been advanced to meet user requirements.
*As of March 2007, according to Shimadzu research
Compact
Only 450 mm wide, the UV-1800 is one of the most compact instruments in its class, allowing installation in tight spaces. In comparison to the UV-1700, the setting space has decreased by about 15%, and the width has narrowed by about 20%.
User friendly
USB memory can be connected directly to the UV-1800. Users can now analyze data on a PC using UVProbe software. In addition, data for spectra and time-course curves can be displayed and saved with commercial spreadsheet software. Printing is possible to printers that support PCL control codes. e.g. HP Business Inkjet 1200 HP Photosmart D5160 Using UVProbe software, provided standard with this instrument, makes it possible to control the UV-1800 with a PC. (A USB cable is required for connection to the PC.)
Security functions
Enabling the UV-1800 security functions makes it possible to restrict functions according to the user level.
Instrument validation functions:
The UV-1800 offers compatibility with 9 JIS items. Wavelength accuracy, wavelength repeatability, resolution, stray light, photometric accuracy, photometric repeatability, baseline flatness, baseline stability, noise level.
Maintenance
The operating time of the deuterium (D2) lamp and the halogen (WI) lamp can be recorded and displayed. Consequently, the expected replacement period of the lamps can be ascertained when performing periodic inspections, minimizing downtime.
General Specifications
Bandwidth | 1 nm (190 to 1,100 nm) |
Depth | 490 mm |
Height | 270 mm |
Width | 450 mm |
Power Requirements | AC100,120,220,230,240 V, 50/60 Hz, 140 VA |
Photometric Range | Absorbance: -4 to 4 Abs, Transmittance: 0% to 400% |
Weight | 15 kg |
Scanning Speed | 3,000 to 2 nm/min |
Light Source | 20-W halogen lamp and deuterium lamp Built-in light source auto position adjustment |
Optics | Double Beam |
Photodetector | Photodiode |
Photometric Noise | Less than 0.00005 Abs (700 nm) |
Sample Compartment | Internal dimensions: 110 (W) x 250 (D) x 115 (H) mm Distance between light beams: 100 mm |
Stray Light | Less than 0.02% at 220 nm (NaI). Less than 0.02% at 340 nm (NaNO2). Less than 1.0% at 198 nm (KCl). |
Additional Specifications
Operating Conditions
Temperature: 15°C to 35°C
Humidity: 30% to 80% (without condensation; 70% max. at 30°C or higher)
Wavelength display: 0.1nm increments
Wavelength setting: 0.1nm increments (1nm increments when setting scanning range)
Wavelength accuracy: ± 0.1 nm at D2 peak 656.1 nm, ± 0.3 nm for entire range
Wavelength repeatability: ± 0.1 nm
Wavelength slew rate: About 6,000 nm/min
Lamp interchange wavelength: Automatic interchange linked to wavelength. The interchange wavelength can be set freely in the range of 295 to 364 nm (0.1nm increments).
Photometric system: Double beam optics
Photometric accuracy:
± 0.002 Abs at 0.5 Abs
± 0.004 Abs at 1.0 Abs
± 0.006 Abs at 2.0 Abs
(measured using NIST930D/NIST1930 or equivalent.)
Photometric repeatability:
Less than ± 0.001 Abs at 0.5 Abs
Less than ± 0.001 Abs at 1 Abs
Less than ± 0.003 Abs at 2 Abs
Baseline stability: Less than 0.0003 Abs/Hr (700 nm, one hour after light source turned ON)
Baseline flatness: Less than ± 0.0006 Abs (1,100 to 190 nm, one hour after light source turned ON)
Output device: USB memory (optional), Data files saved in text format or UVPC format. UVPC-format files can be read directly by UVProbe.
PC compatibility: Provided with UVProbe software. External control possible via USB.